May 2024
Rabea Haddad MD, Edo Birati MD, Hiba Zayyad MD, Nizar Andria MD, Eyal Nachum MD, Erez Kachel MD, Ibrahim Marai MD

Background: Cardiac implantable electronic devices (CIEDs) are increasingly being used; thus, there is an increasing need for transvenous lead extraction (TLE).

Objectives: To summarize our experience with TLE at single referral center in northern Israel.

Methods: The study included all patients who underwent TLE at our center between 2019 and 2022, regardless of the indication.

Results: The cohort included 50 patients. The mean age was 69 ± 10.36 years; 78% were males. A total of 99 electrodes were targeted. The mean number of electrodes was 1.96 (range 1–4) per patient. The time between lead implantation and extraction ranged between 1.1 and 34 years with an average of 8.14 ± 5.71 years (median of 7.5 years). Complete lead removal was achieved in 98% of patients and in 98.99% of leads. The complete procedural success rate as well as the clinical procedural success rate was 96%. The procedural failure rate was 4% (1 patient died 2 days after the index procedure and 1 patient remained with large portion of lead). The indication for TLE was infection in 78% of the cohort group. Powered mechanical sheaths were used in 36 patients (72%), laser sheaths in 27 (54%), and a combination of laser and mechanical sheaths in 16 (32%).

Conclusions: The clinical and procedural success rates of TLE, primarily for CEID-related infection, were high. A combination of laser and mechanical sheaths was needed in one-third of patients.

September 2018
Arwa Younis MD, Anat Wieder MD, Roy Beinart MD, Michael Glikson MD FHRS and Eyal Nof MD

Background: Pacemaker pocket early post-implantation inflammation (EPII) is defined as clinical signs of local erythema without any systemic signs of infection occurring early after implantation. Data on the best treatment regimen for apparent superficial EPII is scarce. 

Objectives: To investigate the prognostic value of medical treatment, rather than extraction surgery, in patients with EPII. 

Methods: Data from 6013 consecutive patients who underwent cardiac implantable electronic device (CIED) implantation or replacement from 2007–2015 were retrospectively analyzed; 40 (0.7%) presented with EPII. Our goal was the absence of major complications and to avoidance of extraction. 

Results: Patients with EPII were initially treated medically. Nineteen (47%) (group A) recovered with antibiotic treatment only. In the other 21 patients (53%) (group B), CIED extraction was performed. Group B had more major complications compared to group A (15 [71%] vs. 0 [0%], P < 0.001). The only significant difference in baseline characteristics was history of non-initial procedure. While 86% of group B patients had a previous non-initial procedure, only 53% of group A patients underwent previous replacement (P < 0.05). In multivariate analysis, previous non-initial procedure was the only predictor for need of extraction at 1 year, hazard ratio 3.5, 95% confidence interval 1.001–11.73, P < 0.05.

Conclusions: Conservative treatment in patients with EPII after non-initial procedure can lead to serious adverse events resulting in the need for extraction. Close follow-up and aggressive treatment should be considered early in the therapeutic course.

January 2016
Amir Givon MD, Natalia Vedernikova MD, David Luria MD, Ori Vatury MD, Rafael Kuperstein MD, Micha S. Feinberg MD, Michael Eldar MD, Michael Glikson MD and Eyal Nof MD

Background: Transvenous lead extraction can lead to tricuspid valve damage. 

Objectives: To assess the incidence, risk factors and clinical outcome of tricuspid regurgitation (TR) following lead extraction.

Methods: We prospectively collected data on patients who underwent lead extraction at the Sheba Medical Center prior to laser use (i.e., before 2012). Echocardiography results before and following the procedure were used to confirm TR worsening, defined as an echocardiographic increase of at least one TR grade. Various clinical and echocardiographic parameters were analyzed as risk factors for TR. Clinical and echocardiographic follow-up was conducted to assess the clinical significance outcome of extraction-induced TR.

Results: Of 152 patients who underwent lead extraction without laser before 2012, 86 (56%) (192 electrodes) had echocardiography results before and within one week following the procedure. New or worsening TR was discovered in 13 patients (15%). Use of mechanical tools and younger age at extraction were found on multivariate analysis to be factors for TR development (P = 0.04 and P = 0.03 respectively). Average follow-up was 22.25 ± 21.34 months (range 8–93). There were no significant differences in the incidence of right-sided heart failure (50% vs. 23%, P = 0.192) or hospitalizations due to heart failure exacerbations (37.5% vs. 11%, P = 0.110). No patient required tricuspid valve repair or replacement. Death rates were similar in the TR and non-TR groups (20% vs. 33%).

Conclusions: TR following lead extraction is not uncommon but does not seem to affect survival or outcomes such as need for valve surgery. Its long-term effects remain to be determined. 

 

Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel