• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Mon, 25.11.24

Search results


April 2024
Dante Antonelli MD, Vladimir Poletaev MD, Vidal Essebag MD, Alexander Feldman MD

Inappropriate implantable cardiac defibrillator (ICD) shock due to electromagnetic interference (EMI) induced by electrocautery is a well-known theoretical association but is rarely reported [1]. We report a case of EMI induced by electrocautery causing inappropriate ICD shock, underlining that, with the use of monopolar cautery, not only the location of the surgery but also electrodispersive pad (EDP) placement may be important to avoid EMI.

June 2023
Dante Antonelli MD, Vladimir Poletaev MD, Alexander Feldman MD

Inappropriate shocks are a serious and still unresolved problem associated with implantable cardioverter defibrillators (ICDs) that have been associated with increased mortality and impairment of quality of life [1] despite advances in device safety. We report a case of electromagnetic interference (EMI) while showering that resulted in an inappropriate ICD discharge.

Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel