• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Mon, 25.11.24

Search results


July 2016
Guy Witberg MD, Ifat Lavi PhD, Hana Vaknin Assa MD, Katia Orvin MD, Abid Assali MD and Ran Kornowski MD FESC FACC

Background: Bioresorbable vascular scaffold (BVS) is a promising technology that potentially offers several advantages over contemporary coronary drug-eluting stents (DES). Crucial to BVS implantation is the correct choice of scaffold size (diameter and length) in order to avoid "geographic miss" in length, provide the maximal support to the vessel wall, and avoid leaving “free-floating” foreign material in the coronary vasculature. 

Objectives: To assess the optimal method for measuring coronary stenosis prior to BVS implantation.

Methods: We compared the performance of two quantitative coronary angiography assessment (QCA) techniques: two dimensional real-time QCA (2D-QCA) and offline 3D QCA (3D-QCA) for the evaluation of coronary lesions in patients enrolled in a multicenter randomized controlled trial of BVS vs. metallic stents, by calculating the weighted kappa value for agreement regarding optimal BVS size with the reference method – CoreLab offline 2D-QCA measurements..In addition, we collected 2 year clinical outcomes (death/myocardial infarction/repeat revascularization/scaffold thrombosis) in BVS-implanted patients.

Results: In 17 patients with available CoreLab data, the weighted kappa for agreement for 3D-QCA was significantly better than for 2D-QCA (0.90, 95%CI 0.72–1.00 vs. 0.439, 95%CI 0.16–0.77). The rate of clinical events at 2 years was low (9.5%).

Conclusions: Initial experience in a small group of carefully selected patients at our institution, suggests that the use of BVS for coronary revascularization is associated with a low rate of adverse events in suitable patients. 3D-QCA may be superior to 2D-QCA analysis in terms of reproducibility, and results in more patients receiving optimal size BVS. 

 

Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel