April 2024
Dante Antonelli MD, Vladimir Poletaev MD, Vidal Essebag MD, Alexander Feldman MD

Inappropriate implantable cardiac defibrillator (ICD) shock due to electromagnetic interference (EMI) induced by electrocautery is a well-known theoretical association but is rarely reported [1]. We report a case of EMI induced by electrocautery causing inappropriate ICD shock, underlining that, with the use of monopolar cautery, not only the location of the surgery but also electrodispersive pad (EDP) placement may be important to avoid EMI.

June 2023
Dante Antonelli MD, Vladimir Poletaev MD, Alexander Feldman MD

Inappropriate shocks are a serious and still unresolved problem associated with implantable cardioverter defibrillators (ICDs) that have been associated with increased mortality and impairment of quality of life [1] despite advances in device safety. We report a case of electromagnetic interference (EMI) while showering that resulted in an inappropriate ICD discharge.

June 2021
Dante Antonelli MD, Alexander Feldman MD, Nahum Adam Freedberg MD, and Yoav Turgeman MD
May 2021
Alexander Feldman MD, Nahum A. Freedberg MD, Dante Antonelli MD, Ehoud Rozner MD, and Yoav Turgeman MD

Background: Patients admitted to the hospital after successful resuscitation from sudden cardiac death (SCD) are treated with therapeutic hypothermia (TH) to facilitate brain preservation. The prognostic significance of J (Osborn) waves (JOW) in the 12 leads electrocardiogram in this setting has not been elucidated as yet.

Objectives: To ascertain retrospectively the prognostic significance of JOW recorded during TH in SCD survivors.

Methods: The study comprised 55 consecutive patients who underwent TH. All patients achieved a core temperature of 33°C at the time of electrocardiogram analysis. We compared 33 patients with JOW to 22 patients without JOW. The endpoints were in-hospital, long-term all-cause mortality, and irreversible anoxic brain injury (IABI).

Results: Patients with JOW compared to patients without JOW were younger (55.1 ± 11.6 vs. 64.5 ± 11.7 years, respectively, P < 0.006), with a lower incidence of hypertension (52% vs. 86%, P < 0.007), diabetes mellitus (15% vs. 50%, P < 0.005), and congestive heart failure (15% vs. 45%, P < 0.013). In-hospital and long-term mortality were significantly higher in patients without JOW (86% vs. 21%, 91% vs. 24%, respectively, P < 0.000001). Among patients without JOW who survived hospitalization, 66.7% presented with IABI versus 7.7% of the patients with JOW (P < 0.0001). In multivariate analysis, the absence of JOW was a significant predictor for poor prognosis.

Conclusions: The absence of J (Osborn) waves on electrocardiograms obtained during TH is associated with poor prognosis among SCD survivors

November 2015
Alexander Feldman MD, Valeria Shaikis MD, Dante Antonelli MD, Nahum Adam Freedberg MD, Malka Yahalom MD DSc and Yoav Turgeman MD