IMAJ | volume 20
Journal 3, March 2018
pages: 147-150
Summary
Background:
Accurate pulse oximetry reading at hospital admission is of utmost importance, mainly for patients presenting with hypoxemia. Nevertheless, there is no accepted or evidence-based protocol for such structured measuring.
Objectives:
To devise and assess a structured protocol intended to increase the accuracy of pulse oximetry measurement at hospital admission.
Methods:
The authors performed a prospective comparison of protocol-based pulse-oximetry measurement with non-protocol based readings in consecutive patients at hospital admission. They also calculated the relative percentage of improvement for each patient (before and after protocol implementation) as a fraction of the change in peripheral capillary oxygen saturation (SpO
2) from 100%.
Results:
A total of 460 patients were recruited during a 6 month period. Implementation of a structured measurement protocol significantly changed saturation values. The SpO
2 values of 24.7% of all study participants increased after protocol implementation (ranging from 1% to 21% increase in SpO
2 values). Among hypoxemic patients (initial SpO
2 < 90%), protocol implementation had a greater impact on final SpO
2 measurements, increasing their median SpO
2 readings by 4% (3–8% interquartile range;
P < 0.05). Among this study population, 50% of the cohort improved by 17% of their overall potential and 25% improved by 50% of their overall improvement potential. As for patients presenting with hypoxemia, the median improvement was 31% of their overall SpO
2 potential.
Conclusions:
Structured, protocol based pulse-oximetry may improve measurement accuracy and reliability. The authors suggest that implementation of such protocols may improve the management of hypoxemic patients.