• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Fri, 22.11.24

Original Articles

IMAJ | volume 19

Journal 9, September 2017
pages: 533-538

Bone Age Assessments by Quantitative Ultrasound (SonicBone) and Hand X-ray Based Methods are Comparable

Summary

Background:

Bone maturation is currently assessed by subjective and automated radiography. 

Objectives:

To evaluate the concordance and reproducibility of a quantitative ultrasound (QUS) based device versus X-ray based methods.

Methods:

The study population comprised 150 children, 76 males, 4–17 years of age. X-ray scans were evaluated according to wrist, carpal and phalanx areas for bone age. QUS was performed by the the BAUS™ device (SonicBone, Rishon Lezion, Israel), using speed-of-sound (SOS) and distance attenuation factor (ATN) in similar areas. Data from 100 subjects were used to establish the device conversion equation, and 50 measurements were assigned to assess inter-modality agreement. 

Results:

BAUS showed high repeatability performance, 0.73% relative standard deviation for SOS and 3.5% for ATN. R2 for the conversion equation, including gender, SOS, and ATN, was 0.80 for all methods (P < 0.001). There was no significant bias in bone age assessments.

Conclusions:

Bone age assessment by SonicBone is comparable to the assessment by X-ray based methods. 
 

Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
© All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel