IMAJ | volume 18
Journal 10, October 2016
pages: 609-612
Summary
Background:
Automated breast volumetric sonography (ABVS) is a new technology with various possible applications.
Objectives:
To compare ABVS and breast magnetic resonance imaging (MRI) in the surveillance of women with BRCA1/2 gene mutation carriers.
Methods:
We conducted a prospective study in Jewish female BRCA1/2 mutation carriers who underwent breast MRI and ABVS. The results of both exams performed 6 months apart or less, and relevant clinical data, were reviewed. The BIRADS results were divided into three subgroups according to subsequent expected management: BIRADS 1-2 (normal study), BIRADS 3 (probably benign finding), and BIRADS 4 and 5 (suspicious findings). BIRADS 0 and 6 scores were excluded from the study. Distribution of ABVS and MRI BIRADS scores were compared using McNemar's test, and concordance was calculated using the Cohen kappa test.
Results:
Overall, 68 women, 40 BRCA1 and 28 BRCA2 mutation carriers, age range 26–69 (mean 44.55 ± 12.1 years), underwent 79 paired ABVS and MRI examinations. McNemar's test calculations showed no significant difference between MRI and ABVS BIRADS score distribution. Cohen’s kappa test resulted in k = 0.158, an agreement that can be described as only "slight agreement" between both modalities. Of 14 discordant cases there was one cancer, revealed by MRI and not by ABVS performed 6 months prior to MRI.
Conclusions:
ABVS showed slight agreement with MRI in BRCA1/2 mutation carriers. These preliminary results on a small group of healthy high risk patients suggest that the diagnostic abilities of ABVS are inferior to MRI. Further studies encompassing larger groups are needed.