• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Fri, 22.11.24

Original Articles

IMAJ | volume 15

Journal 9, September 2013
pages: 538-544

Long-Term Clinical and Echocardiographic Follow-Up of the Freestyle Stentless Aortic Bioprosthesis: the Tel Aviv Medical Center Experience

    Summary

    Background: Stentless aortic bioprostheses were designed to provide improved hemodynamic performance and potentially better survival.

    Objectives: To report the outcomes of patients after aortic valve replacement with the Freestyle® stentless bioprosthesis in the Tel Aviv Medical Center followed for ≤ 15 years.

    Methods and Results: Between 1997 and 2011, 268 patients underwent primary aortic valve replacement with a Freestyle bioprosthesis, 211 (79%) of them in the sub-coronary position. Mean age, Charlson comorbidity index and Euro-score were 71.0 ± 9.2 years, 4.2 ± 1.5 and 10.2 ± 11 respectively, and 156 (58%) were male. Peak and mean trans-aortic gradient decreased significantly (75.0 ± 29.1 vs. 22.8 ± 9.6 mmHg, P < 0.0001; and 43.4 ± 17.2 vs. 12.1 ± 5.4 mmHg, P < 0.0001 respectively) in 3 months of follow-up. Mean overall follow-up was 4.9 ± 3.1 years and was complete in all patients. In-hospital mortality was 4.1% (n=11) but differed significantly between the first 100 patients operated before 2006 and the last 168 patients operated after January 2006 (8 vs. 3 patients, 8.0% vs. 1.8%, P = 0.01). Overall, 5 and 10 year survival rates were 85 ± 2.5% and 57.2 ± 5.7%, respectively. Five year survival was markedly improved in patients operated after January 2006 compared to those operated in the early years of the experience (92.3 ± 2.3% vs. 76.0 ± 4.4%, P = 0.0009). All the 21 octogenarians operated after January 2006 survived surgery, with excellent 5 year survival (85.1 ± 7.9%). Six patients required reoperation during follow-up: structural valve deterioration in five and endocarditis in one.

    Conclusions: Aortic valve replacement with the Freestyle bioprosthesis provides good long-term hemodynamic and clinical outcomes, even in octogenarians. Valve calcification is the major (and rare) mode of valve deterioration leading to reoperation in these patients. 

    Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
    The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
    © All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy

    2 Twin Towers, 35 Jabotinsky, POB 4292, Ramat Gan 5251108 Israel