• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Fri, 22.11.24

Search results


August 2013
A. Segev, D. Spiegelstein, P. Fefer, A. Shinfeld, I. Hay, E. Raanani and V. Guetta

Background: Trans-catheter aortic valve implantation (TAVI) has emerged as a novel therapeutic approach for patients with severe tricuspid aortic stenosis (AS) not suitable for aortic valve replacement.

Objectives: To describe our initial single-center experience with TAVI in patients with "off-label" indications.

Methods: Between August 2008 and December 2011 we performed TAVI in 186 patients using trans-femoral, trans-axillary, trans-apical and trans-aortic approaches. In 11 patients (5.9%) TAVI was undertaken due to: a) pure severe aortic regurgitation (AR) (n=2), b) prosthetic aortic valve (AV) failure (n=5), c) bicuspid AV stenosis (n=2), and d) prosthetic valve severe mitral regurgitation (MR) (n=2).

Results: Implantation was successful in all: six patients received a CoreValve and five patients an Edwards-Sapien valve. In-hospital mortality was 0%. Valve hemodynamics and function were excellent in all patients except for one who received an Edwards-Sapien that was inside a Mitroflow prosthetic AV and led to consistently high trans-aortic gradients. No significant residual regurgitation in AR and MR cases was observed.
Conclusions: TAVI is a good alternative to surgical AV replacement in high risk or inoperable patients with severe AS. TAVI for non-classical indications such as pure AR, bicuspid AV, and failed prosthetic aortic and mitral valves is feasible and safe and may be considered in selected patients. 

April 2007
Y. Shapira, D. E. Weisenberg, M. Vaturi, E. Sharoni, E. Raanani, G. Sahar, B. A. Vidne, A. Battler and A. Sagie

Backgound: The use of intraoperative transesophageal echocardiogram in patients with infective endocarditis is usually reserved for cases of inadequate preoperative testing or suspected extension to perivalvular tissue.

Objectives: To explore the impact of routine intraoperative TEE[1] in patients with infective endocarditis.

Methods: The impact of intraoperative TEE on the operative plan, anatomic-physiologic results, and hemodynamic assessment or de-airing was analyzed in 59 patients (38 males, 21 females, mean age 57.7 ± 16.8 years, range 20–82) operated for active infective endocarditis over 56 months.

Results: Immediate pre-pump echocardiography was available in 52 operations (86.7%), and changed the operative plan in 6 of them (11.5%). Immediate post-pump study was available in 59 patients (98.3%) and accounted for second pump-run in 6 (10.2%): perivalvular leak (3 cases), and immobilized leaflet, significant mitral regurgitation following vegetectomy, and failing right ventricle requiring addition of vein graft (1 case each). Prolonged de-airing was necessary in 6 patients (10.2%). In 5 patients (8.5%) the postoperative study aided in the evaluation and treatment of difficult weaning from the cardiopulmonary bypass pump. In 21 patients (35.6%) the application of intraoperative TEE affected at least one of the four pre-specified parameters.
Conclusions: Intraoperative TEE has an important role in surgery for infective endocarditis and should be routinely implemented







[1] TEE = transesophageal echocardiogram