• IMA sites
  • IMAJ services
  • IMA journals
  • Follow us
  • Alternate Text Alternate Text
עמוד בית
Thu, 21.11.24

Original Articles

IMAJ | volume 18

Journal 2, February 2016
pages: 104-107

Retrievable Inferior Vena Cava Filters: Indications, Indweling Time, Retrieval, Success, and Complication Rates

    Summary

    Background:

    Various vena cava filters (VCF) are designed with the ability to be retrieved percutaneously. Yet, despite this option most of them remain in the inferior vena cava (IVC). 

    Objectives:

    To report our experience in the placement and retrieval of three different types of VCFs, and to compare the indications for their insertion and retrieval as reported in the literature.

    Methods:

    During a 5 year period three types of retrievable VCF (ALN, OptEase, and Celect) were inserted in 306 patients at the Rabin Medical Center (Beilinson and Hasharon hospitals). Indications, retrieval rates, median time to retrieval, success and complications rate were viewed and assessed in the three groups of filter types and were compared with the data of similar studies in the literature.

    Results:

    Of the 306 VCFs inserted, 31 (10.1%) were retrieved with equal distribution in the three groups. In most patients the reason for filter insertion was venous thromboembolic events (VTE) and contraindications to anticoagulant therapy. Mean age was 68.38 ± 17.5 years (range 18–99) and was noted to be significantly higher compared to similar studies (53–56 years) (P < 0.0001). Multi-trauma patients were significantly older (71.11 ± 14.99 years) than post-pulmonary embolism patients (48.03 ± 20.98 years, P < 0.0001) and patients with preventive indication (26.00 ± 11.31, P < 0.0001). The mean indwelling time was 100.6 ± 103.399 days. Our results are comparable with the results of other studies, and there was no difference in percentage of retrieval or complications between patients in each of the three groups. 

    Conclusions:

    In 1 of 10 patients filters should be removed after an average of 3.5 months. All three IVC filter types used are safe to insert and retrieve.

     

    Legal Disclaimer: The information contained in this website is provided for informational purposes only, and should not be construed as legal or medical advice on any matter.
    The IMA is not responsible for and expressly disclaims liability for damages of any kind arising from the use of or reliance on information contained within the site.
    © All rights to information on this site are reserved and are the property of the Israeli Medical Association. Privacy policy